• DocumentCode
    943360
  • Title

    Phase-Controlled Circular Array Heating Equipment for Deep-Seated Tumors: Preliminary Experiments

  • Author

    Sato, Gentei ; Shibata, Chokichiro ; Sekimukai, Shuichi ; Wakabayashi, Hiroshi ; Mitsuka, Kaoru ; Giga, Kenjiro

  • Volume
    34
  • Issue
    5
  • fYear
    1986
  • fDate
    5/1/1986 12:00:00 AM
  • Firstpage
    520
  • Lastpage
    525
  • Abstract
    This paper presents some preliminary results on the development of a circular phased-array equipment for heating deep-seated tumors. It is shown that radiators having sharp directivity are needed to realize excellent focusing of SAR. Moreover, moment method calculations indicate that linearly polarized helical radiators inversely wound with double wires possess the desired directivity and the near field pattern. Field patterns were measured in phantom (plastic cylinders containing saline water) with a circular array formed by four pairs of azimuthally positioned radiators. Excellent focusing of SAR was observed. The ratio of valley to peak value was 0.74. Steering of SAR maximum by phase control was observed and the distance of movement coincided with that estimated from phase variation.
  • Keywords
    Heating; Imaging phantoms; Moment methods; Neoplasms; Phased arrays; Plastics; Polarization; Position measurement; Wires; Wounds;
  • fLanguage
    English
  • Journal_Title
    Microwave Theory and Techniques, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9480
  • Type

    jour

  • DOI
    10.1109/TMTT.1986.1133386
  • Filename
    1133386