Abstract :
A whole-body positron-emission tomography (PET) machine based on a novel type of large-area sodium iodide (NaI) detector is being designed. Monte Carlo computer simulations have been used to compare some of the performance parameters of a tomograph based on the detectors to similar parameters of conventional small-crystal machines. Three different prototypes have been constructed, and many tests have been performed, including measurements of transverse spatial resolution, depth-of-interaction resolution, energy resolution, time resolution, and high-counting-rate capabilities. The simulation and measurement results are presented and discussed.<>
Keywords :
biomedical equipment; computerised tomography; radioisotope scanning and imaging; Monte Carlo computer simulations; NaI; depth-of-interaction resolution; energy resolution; high-counting-rate capabilities; time resolution; transverse spatial resolution; whole body PET machine; whole-body positron-emission tomography; Computer simulation; Detectors; Energy resolution; Monte Carlo methods; Performance evaluation; Prototypes; Spatial resolution; Testing; Tomography; Whole-body PET;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/23.12811
Filename :
12811
Link To Document :
بازگشت