  • Abstract
    Several planar coil designs, with integral resonating capacitors, are reported using large-area sputtered YBCO films. One design utilizes a distributed interdigital capacitor, while a second exploits capacitive coupling of two films on opposite sides of a dielectric. Devices of the first design made on LaAlO/sub 3/ have achieved Q>10/sup 4/ at 18.5 MHz, possibly limited by the loss tangent of this substrate. Identical coils made with silver films have Q=10. With the second design, Q>10/sup 4/ was achieved at the even lower frequency of 5.7 MHz.<>
  • Keywords
    Q-factor; barium compounds; biomedical NMR; biomedical electronics; biomedical equipment; coils; high-temperature superconductors; probes; superconducting magnets; superconducting thin films; thin film devices; yttrium compounds; 18.5 MHz; 5.7 MHz; LaAlO/sub 3/; MRI; YBa/sub 2/Cu/sub 3/O/sub 7/-LaAlO/sub 3/; capacitive coupling; distributed interdigital capacitor; integral resonating capacitors; large area sputtered films; loss tangent; magnetic-resonance imaging; planar coil designs; probe coils; Capacitors; Coils; Dielectric substrates; Dielectric thin films; High temperature superconductors; Magnetic films; Probes; Silver; Sputtering; Yttrium barium copper oxide;
  • fLanguage
    English
  • Journal_Title
    Applied Superconductivity, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    1051-8223
  • Type

    jour

  • DOI
    10.1109/77.233951
  • Filename
    233951