  • Title

    Post-injection transmission attenuation measurements for PET

  • Author

    Daube-Witherspoon, Margaret E. ; Carson, Richard E. ; Green, Michael V.

  • Author_Institution
    Nat. Inst. of Health, Bethesda, MD, USA
  • Volume
    35
  • Issue
    1
  • fYear
    1988
  • Firstpage
    757
  • Lastpage
    761
  • Abstract
    The acquisition of positron-emission-tomography (PET) transmission information after tracer injection using ring and rotating pin transmission sources is discussed. A combined transmission/emission scan was acquired, followed by an emission scan used to substract the emission counts from the transmission/emission data. The ratio of emission count rate for brain scans to transmission count rate is 50-100% for a 5-mCi ring source and less than 5% for a 5-mCi pin source. Windowing of the sinogram, which rejects most random and scattered coincidences, also eliminates most emission counts. The magnitude and effects of residual random and scattered coincidences as well as increases in variability from transmission/emission scans were studied. The results of combined transmission/emission scans for a high-contrast emission source distribution using ring and pin sources are also described.<>
  • Keywords
    computerised tomography; radioisotope scanning and imaging; 5*10/sup -3/ Ci; PET; acquisition; brain scans; combined transmission/emission scan; positron-emission-tomography; ring; rotating pin transmission sources; sinogram; tracer injection; transmission information; Attenuation measurement; Error correction; Imaging phantoms; Nuclear medicine; Performance evaluation; Pollution measurement; Positron emission tomography; Rotation measurement; Scattering; Time measurement;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.12827
  • Filename
    12827