  • Abstract
    Experiments were carried out for studying the resolution of 4-in (10-cm)×4-in (10-cm)×16-in (40-cm) NaI(Tl) detectors used in high-sensitivity airborne gamma-ray spectrometric surveys. Change of resolution with source-detector geometry emphasized the inadequacy of existing specifications for resolution in explaining the detector response. An additional specification for resolution measurement, namely, overall resolution, is proposed. Measurement and applications of overall resolution of the detector are discussed
  • Keywords
    cosmic ray apparatus; gamma-ray detection and measurement; scintillation counters; sodium compounds; thallium; NaI:Tl; detector response; high-sensitivity airborne gamma-ray spectrometric surveys; resolution measurement; source-detector geometry; Crystals; Energy resolution; Gamma ray detection; Gamma ray detectors; Helium; Preamplifiers; Radiometry; Sensor arrays; Spectroscopy; Voltage;