Improvement of SOI microdosimeter performance using pulse-shape discrimination techniques
Author :
Cornelius, Iwan ; Rosenfeld, Anatoly ; Siegele, Rainer ; Cohen, David D.
Author_Institution :
Centre for Med. Radiat. Phys., Wollongong Univ., NSW, Australia
Volume :
49
Issue :
6
fYear :
2002
fDate :
12/1/2002 12:00:00 AM
Firstpage :
2805
Lastpage :
2809
Abstract :
The timing properties of a silicon-on-insulator microdosimeter for medical and space applications have been studied using an ion microprobe. These measurements were used with a pulse-shape discrimination technique to render the microdosimeter insensitive to ion strikes outside the ideal sensitive volume. These improvements have resulted in a microdosimeter with a cubic sensitive volume with dimensions 10×10×10 μm3, a charge collection spectrum close to Gaussian for a monoenergetic source, and a decreased sensitivity to radiation damage.
Keywords :
dosimeters; ion beam effects; silicon radiation detectors; silicon-on-insulator; Gaussian distribution; SOI microdosimeter; Si; charge collection spectrum; ion microprobe; ion strike; pulse-shape discrimination; radiation damage; sensitive volume; timing properties; Aerospace electronics; Australia; Diodes; Energy measurement; Medical treatment; Physics; Pulse measurements; Radiation detectors; Silicon; Volume measurement;