• DocumentCode
    953473
  • Title

    Arterial Vulnerable Plaque Characterization Using Ultrasound-Induced Thermal Strain Imaging (TSI)

  • Author

    Kim, Kang ; Huang, Sheng-Wen ; Hall, Timothy L. ; Witte, Russell S. ; Chenevert, Thomas L. ; O´Donnell, Matthew

  • Author_Institution
    Michigan Univ., Ann Arbor
  • Volume
    55
  • Issue
    1
  • fYear
    2008
  • Firstpage
    171
  • Lastpage
    180
  • Abstract
    Thermal strain imaging (TSI) is demonstrated in two model systems mimicking two potential clinical applications. First, a custom ultrasound (US) microscope produced high-resolution TSI images of an excised porcine coronary artery. Samples were placed in a temperature-controlled water chamber and scanned transversely and longitudinally. Phase-sensitive, correlation-based speckle tracking was applied to map the spatial distribution of temporal strain across the sample. TSI differentiated fatty tissue from water-based arterial wall and muscle with high contrast and a spatial resolution of 60 mum for a 50-MHz transducer. Both transverse and longitudinal TSI images compared well with B-scans of arterial wall structures, including intima, media, adventitia, and overlying fatty tissue. A second model system was used to test the hypothesis that US can produce the heating pattern required for TSI of internal structures. A 2-D phased array with independent drive electronics was combined with a conventional US scanner (iU22, Philips, Bothell, WA) for these studies. This 513-element array, originally designed for the US therapy, acted as the US heat source. To quantify the temporal strain induced by this system, TSI was performed on a homogeneous rubber phantom. TSI temperature estimates were within 3% error for a 3.2degC temperature rise produced within 2 s using a specially designed beamformer and pulse sequencer. The system was then used to produce TSI scanning of an excised kidney containing an intact piece of fat below the collecting system. These images were validated using an magnetic resonance imaging (MRI) pulse sequence designed for lipid quantification. TSI scans matched well MRI scans and histology both anatomically and quantitatively. Finally, to test the potential of US-induced TSI for a significant clinical problem, images were obtained on an excised canine aorta with fatty tissue inside the lumen. Both longitudinal and transversal TSI agreed well with anatomy. T- - hese in vitro results demonstrate the potential of high-resolution US-induced TSI with a small temperature change (<1degC) for plaque characterization.
  • Keywords
    biomedical ultrasonics; biothermics; cardiovascular system; muscle; arterial wall structures; coronary artery; frequency 50 MHz; histology; magnetic resonance imaging; phase sensitive speckle tracking; plaque characterization; rubber phantom; ultrasound microscope; ultrasound-induced thermal strain imaging; Arteries; Capacitive sensors; High-resolution imaging; Magnetic resonance imaging; Microscopy; Muscles; Phased arrays; Speckle; Temperature; Ultrasonic imaging; Cardiovascular Plaque Characterization; Thermal Strain Imaging; Thermal strain imaging; Ultrasound Induced Thermal Strain Imaging; cardiovascular plaque characterization; ultrasound-induced thermal strain imaging; Animals; Coronary Artery Disease; Coronary Vessels; Echocardiography; Elasticity; Feasibility Studies; Image Interpretation, Computer-Assisted; Microscopy, Acoustic; Stress, Mechanical; Swine; Thermography;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.2007.900565
  • Filename
    4360058