Title :
Development of an Implantable Pulse Oximeter
Author :
Reichelt, Stephan ; Fiala, Jens ; Werber, Armin ; Förster, Katharina ; Heilmann, Claudia ; Klemm, Rolf ; Zappe, Hans
Author_Institution :
Univ. of Freiburg, Freiburg
Abstract :
A long-term implantable photoplethysmographic sensor system is proposed. The system employs an elastic cuff which is directly wrapped around an arterial blood vessel. The optically transparent cuff is equipped with light emitting diodes and a photo transistor including the technology of pulse oximetry. The sensor will permit real-time, continuous monitoring of important vital parameters such as arterial blood oxygen saturation and pulse rate over a long-term period in vivo. We emphasize on the specific requirements for design and instrumentation of the implantable sensor and discuss first in vitro data acquired with that new photonics-based sensor.
Keywords :
blood vessels; cardiology; data acquisition; light emitting diodes; optical sensors; oximetry; patient monitoring; phototransistors; plethysmography; arterial blood oxygen saturation; blood vessel; elastic cuff; implantable pulse oximeter; in vitro data acquisition; light emitting diodes; long-term photoplethysmographic sensor system; optically transparent cuff; photo transistor; photonics-based sensor; real-time continuous monitoring; Blood vessels; Condition monitoring; In vivo; Light emitting diodes; Optical pulses; Optical saturation; Optical sensors; Optoelectronic and photonic sensors; Sensor systems; Stimulated emission; Implant sensor; Pulse oximetry; implant sesor; long-term monitoring; oxygen saturation; physiological sensing; pulse oximetry; Diagnosis, Computer-Assisted; Equipment Design; Equipment Failure Analysis; Lighting; Monitoring, Ambulatory; Oximetry; Prostheses and Implants; Reproducibility of Results; Semiconductors; Sensitivity and Specificity; Signal Processing, Computer-Assisted; Transducers;
Journal_Title :
Biomedical Engineering, IEEE Transactions on
DOI :
10.1109/TBME.2007.902242