Title :
New a.d. convertor with special application to medium precision platinum resistance thermometry
Author_Institution :
CSIRO, Division of Applied Physics, Sydney, Australia
Abstract :
Platinum resistance thermometers mounted on ceramic formers have a potential uncertainty, after calibration, of 0.01°C over a range exceeding ¿30 to +250°C. To achieve this requires resistance measurements of higher precision than is conveniently available from commercial digital instruments. A new time-interval type of a.d. convertor has been developed for this purpose and incorporated in a simple digital resistance meter with a full scale reading of 300.000 ¿ and an uncertainty of 1 m¿.
Keywords :
analogue-digital conversion; digital instrumentation; electric variables measurement; platinum; resistance thermometers; AD convertor; Pt resistance thermometry; ceramic formers; digital resistance meter; resistance measurements;
Journal_Title :
Electronics Letters
DOI :
10.1049/el:19800401