DocumentCode
972232
Title
Medical Navigation System for Otologic Surgery Based on Hybrid Registration and Virtual Intraoperative Computed Tomography
Author
Hong, Jaesung ; Matsumoto, Nozomu ; Ouchida, Riichi ; Komune, Shizuo ; Hashizume, Makoto
Author_Institution
Innovation Center for Med. Redox Navig., Kyushu Univ., Fukuoka
Volume
56
Issue
2
fYear
2009
Firstpage
426
Lastpage
432
Abstract
An image-guided surgical system for otologic surgery was developed and clinically evaluated. With reliable hybrid registration, real-time patient movement compensation and virtual intraoperative computed tomography imaging have been originally proposed. In contrast to the commercially available systems that mainly use 2-D images for pointing probes, in this system, the surgical drill position is navigated and displayed in the 3-D space with real-time surface rendering. In a temporal bone model study, the navigation accuracy was 1.12 plusmn 0.09 mm with regard to the target registration error. Initial clinical evaluation of the proposed method was performed in five cochlea implantation surgeries. Accurate insertion of the electrodes into the cochlea was achieved, and the facial nerve was protected from injury in all surgeries. The proposed method could be applied to various surgeries for accurate targeting and protection of critical organs.
Keywords
bone; cochlear implants; computerised tomography; image registration; medical image processing; real-time systems; surgery; cochlea implantation surgery; hybrid registration; image-guided surgical system; navigation accuracy; otologic surgery; real-time patient movement compensation; real-time surface rendering; surgical drill position; target registration error; temporal bone model; virtual intraoperative computed tomography imaging; Biomedical imaging; Bones; Computed tomography; Navigation; Oncological surgery; Performance evaluation; Probes; Real time systems; Rendering (computer graphics); Surge protection; Image-guided surgery; otologic surgery; surgical navigation; virtual computed tomography (CT); Aged; Cochlear Implantation; Equipment Design; Equipment Failure Analysis; Female; Humans; Image Processing, Computer-Assisted; Male; Middle Aged; Models, Biological; Movement; Otologic Surgical Procedures; Phantoms, Imaging; Surgery, Computer-Assisted; Temporal Bone; Tomography, X-Ray Computed;
fLanguage
English
Journal_Title
Biomedical Engineering, IEEE Transactions on
Publisher
ieee
ISSN
0018-9294
Type
jour
DOI
10.1109/TBME.2008.2008168
Filename
4663632
Link To Document