• DocumentCode
    973477
  • Title

    Voltage ratio technique used with the national RF standard attenuation measuring equipment at the PTB

  • Author

    Rühle, Wolfgang ; Stumpe, Dieter

  • Author_Institution
    Physikalisch Tech. Bundesanstalt, Braunschweig, Germany
  • Volume
    42
  • Issue
    6
  • fYear
    1993
  • fDate
    12/1/1993 12:00:00 AM
  • Firstpage
    1009
  • Lastpage
    1013
  • Abstract
    With a new measuring system, the attenuation range of the national RF standard attenuation measuring equipment at the Physikalisch-Technische Bundesanstalt (PTB), Germany, has been extended to 60 db in the waveguide bands R 220 and R 320. Most of the system is automated to shorten the calibration and measurement time. The total uncertainty of measurement is estimated to be 0.0027 dB per 10 dB attenuation
  • Keywords
    attenuation measurement; calibration; measurement standards; microwave measurement; 18 to 40 GHz; 60 dB; Germany; PTB; Physikalisch-Technische Bundesanstalt; R 220; R 320; attenuation measurement; calibration; measurement time; national RF standard; uncertainty; waveguide bands; Attenuation measurement; Calibration; Couplers; Dynamic range; Frequency measurement; Frequency synthesizers; Measurement standards; Power measurement; Radio frequency; Voltage;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.245654
  • Filename
    245654