• DocumentCode
    975399
  • Title

    The effect of shaking on magnetic shields

  • Author

    Kelhä, Väinö O. ; Peltonen, R. ; Rantala, Börje

  • Author_Institution
    Technical Research Centre of Finland, Metallimiehenkuja, Espoo, Finland
  • Volume
    16
  • Issue
    4
  • fYear
    1980
  • fDate
    7/1/1980 12:00:00 AM
  • Firstpage
    575
  • Lastpage
    578
  • Abstract
    The increase of the shielding factor due to shaking was measured in a scale model for a magnetically shielded room. The increase was found to be 7 dB for a single-layer square cylinder biased by the Earth´s magnetic field. The shielding factor of a large-volume three-layer Mumetal®room was estimated to increase by a factor of 30, thus confirming the feasibility of shaking in magnetic shields. The shaking parameters, amplitude, and frequency are not critical according to the experiments. Winding the shaking coils along the edges of the cubic shield leads to minimum disturbances inside the cube, and the winding can also be applied to demagnetize the shield by an alternating field of 25 A/m, 50 Hz. The relative incremental permeability of Mumetal was studied as a function of the shaking and biasing fields. The permeability was found to increase considerably by shaking and by decreasing the biasing field. With zero biasing and with shaking field of Hs= 5 A/m root mean square (rms), 50 Hz, the permeability reached its maximum value of 89 000, which is sevenfold the value without shaking.
  • Keywords
    Magnetic shielding; Coils; Frequency; Instruments; Laboratories; Magnetic field measurement; Magnetic materials; Magnetic shielding; Permeability; Temperature; Toroidal magnetic fields;
  • fLanguage
    English
  • Journal_Title
    Magnetics, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9464
  • Type

    jour

  • DOI
    10.1109/TMAG.1980.1060655
  • Filename
    1060655