DocumentCode
978255
Title
Preliminary investigations on a new disposable potentiometric biosensor for uric acid
Author
Liao, Cheng Wei ; Chou, Jung Chuan ; Sun, Tai Ping ; Hsiung, Shen Kan ; Hsieh, Jui Hsiang
Author_Institution
Inst. of Biomed. Eng., Chung Yuan Christian Univ., Taiwan
Volume
53
Issue
7
fYear
2006
fDate
7/1/2006 12:00:00 AM
Firstpage
1401
Lastpage
1408
Abstract
In this paper, uricase, catalase, and electron mediator were coimmobilized on the surface of the tin oxide (SnO2)/indium tin oxide (ITO) glass, to develop a disposable potentiometric uric acid biosensor. The SnO2 / ITO glass was employed as a pH sensor, fabricated by sputtering SnO2 thin films on the ITO glass. 3-Glycidyloxypropyltrimethoxysilane (GPTS) was utilized to immobilize uricase, catalase and the electron mediator (ferrocenecarboxylic acid, FcA) on the sensing window. The experimental results reveal that the optimal weight ratio of uricase, FcA to catalase (CAT) is 4:1:2. The sensor responds linearly between 2 mg/dl and 7 mg/dl at pH 7.5, in 20 mM of test solution, with a correlation coefficient of 0.99213. Accordingly, no significant interference was observed when interfering substances, glucose, urea and ascorbic acid, were added to the uric acid solution. Moreover, the recorded voltage was relatively constant during the first 28 days of measurement. Consequently, a potentiometric uric acid biosensor was realized with the advantages of low cost and simple fabrication.
Keywords
bioelectric potentials; biosensors; chemical sensors; enzymes; molecular biophysics; patient diagnosis; potentiometers; 3-glycidyloxypropyltrimethoxysilane; SnO/sub 2/-In/sub 2-x/Sn/sub x/O/sub 3-y/; catalase; disposable potentiometric biosensor; electron mediator; ferrocenecarboxylic acid; pH sensor; tin oxide/indium tin oxide glass; uric acid; uricase; Biosensors; Electrons; Glass; Indium tin oxide; Interference; Sputtering; Sugar; Testing; Thin film sensors; Windows; 3-glycidyloxypropyltrimethoxysilane (GPTS); Electron mediator; pH sensor; uricase; Biosensing Techniques; Electrochemistry; Equipment Design; Equipment Failure Analysis; Hydrogen-Ion Concentration; Microelectrodes; Pilot Projects; Reproducibility of Results; Sensitivity and Specificity; Urate Oxidase; Uric Acid; Urinalysis;
fLanguage
English
Journal_Title
Biomedical Engineering, IEEE Transactions on
Publisher
ieee
ISSN
0018-9294
Type
jour
DOI
10.1109/TBME.2006.875720
Filename
1643409
Link To Document