• DocumentCode
    979055
  • Title

    Efficient electrode spacing for examining spatial organization during ventricular fibrillation

  • Author

    Bayly, Philip V. ; Johnson, Eric E. ; Idriss, Salim F. ; Ideker, Raymond E. ; Smith, William M.

  • Author_Institution
    Sch. of Eng., Duke Univ., Durham, NC, USA
  • Volume
    40
  • Issue
    10
  • fYear
    1993
  • Firstpage
    1060
  • Lastpage
    1066
  • Abstract
    Spatial organization has been observed during episodes of ventricular fibrillation (VF) by recording epicardial unipolar electrograms on a grid of electrodes. In such studies, the choice of spacing between electrodes is an important decision, affecting the resolution and the size of the domain to be studied. A basic tenet of sampling theory, the Nyquist criterion, states that an electrode spacing smaller than half the smallest significant wavelength is required to capture the important details of a spatially sampled process. The authors suggest a method to choose a practical interelectrode spacing by examining wavenumber power spectra of high-resolution VF data recorded from a square 11*11 array of electrodes spaced 0.28 mm apart. The plaque was sutured on the epicardium near the left ventricular apex in 7 anesthetized pigs. VF was induced with AC simulation. Unipolar extracellular electrograms were simultaneously recorded from each channel for 2 s after the induction of VF. Each signal was sampled in time at 1000 Hz. Wavenumber power spectra were calculated for 100 ms segments using the zero-delay wavenumber spectrum method, for a total of 140 power spectra. All spectra had dominant peaks at the origin and fell off rapidly with increasing wavenumber (decreasing wavelength). In all the spectra, every wavelength shorter than 1.4 mm contributed insignificant power. Furthermore, in 134 of 140 spectra (96%), insignificant power levels were associated with every wavelength shorter than 2.8 mm. These results suggest that, for unipolar extracellular electrodes, an intersensor spacing on the order of 1 mm is appropriate to study organization during early VF.
  • Keywords
    biological techniques and instruments; electrocardiography; electrodes; 100 ms; 1000 Hz; 2 s; AC simulation; Nyquist criterion; anesthetized pigs; efficient electrode spacing; electrodes grid; epicardial unipolar electrograms; epicardium; intersensor spacing; left ventricular apex; plaque; power spectra; spatial organization examination; unipolar extracellular electrodes; ventricular fibrillation; zero-delay wavenumber spectrum method; Biomedical engineering; Biomedical imaging; Continuous time systems; Electrodes; Extracellular; Fibrillation; Frequency; Heart; Sampling methods; Spatial resolution; Animals; Electrocardiography; Electrodes; Swine; Ventricular Fibrillation;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/10.247805
  • Filename
    247805