DocumentCode
981279
Title
Wavefront amplitude distortion and image sidelobe levels. II. In vivo experiments
Author
Zhu, Qing ; Steinberg, Bernard D. ; Arenson, Ronald L.
Author_Institution
Moore School of Electr. Eng., Pennsylvania Univ., Philadelphia, PA, USA
Volume
40
Issue
6
fYear
1993
Firstpage
754
Lastpage
762
Abstract
For part I, see ibid., vol.40, no.6, p.747-753 (1993). In vivo measurements of the rise of the sidelobe level in a single-source image obtained through the female breast as a function of the distortion of the wavefront amplitude are described. The measured sidelobe levels are the average sidelobe floor (ASF) and the peak sidelobe level (PSL). The ASF is shown to be proportional to the variance of the modulus of the wavefront normalized to the square of its mean value, with a proportionality constant close to the value predicted by theory. The PSL similarly increases linearly. The average ratio of PSL to ASF is 5 (7 dB).<>
Keywords
acoustic imaging; biomedical ultrasonics; medical image processing; ASF; PSL; average sidelobe floor; female breast; image sidelobe levels; in-vivo measurement; peak sidelobe level; single-source image; ultrasonic propagation; wavefront amplitude distortion; wavefront modulus variance; Breast; Distortion measurement; Focusing; In vivo; Phase distortion; Phase measurement; Pulse measurements; Radiology; Ultrasonic transducer arrays; Ultrasonic variables measurement;
fLanguage
English
Journal_Title
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher
ieee
ISSN
0885-3010
Type
jour
DOI
10.1109/58.248220
Filename
248220
Link To Document