DocumentCode
983176
Title
Improving the traceability of coaxial impedance measurements at lower RF in the UK
Author
Ridler, N.M. ; Medley, J.C.
Volume
143
Issue
4
fYear
1996
fDate
7/1/1996 12:00:00 AM
Firstpage
241
Lastpage
246
Abstract
The authors report on a measurement service introduced recently by NPL at Malvern to provide traceability for impedance, and other complex reflection parameters, for coaxial line terminations fitted with APC-7 connectors at frequencies below 50 MHz. The new service is a result of several improvements made to an existing measurement system. These improvements, including the use of robust statistical techniques to reduce the uncertainty of measurement, are discussed and typical results produced by the enhanced system are presented
Keywords
calibration; coaxial cables; electric connectors; electric impedance measurement; electromagnetic wave reflection; measurement standards; statistical analysis; 100 kHz to 50 MHz; APC-7 connectors; NPL; coaxial impedance measurements; coaxial line terminations; complex reflection parameters; low frequency calibration; lower RF range; measurement service; measurement uncertainty; robust statistical techniques; traceability;
fLanguage
English
Journal_Title
Science, Measurement and Technology, IEE Proceedings -
Publisher
iet
ISSN
1350-2344
Type
jour
DOI
10.1049/ip-smt:19960524
Filename
503751
Link To Document