DocumentCode :
984073
Title :
Micro-Magnetocardiography System With a Single-Chip SQUID Magnetometer Array for QT Analysis and Diagnosis of Myocardial Injury in Small Animals
Author :
Komamura, K. ; Adachi, Y. ; Miyamoto, M. ; Kawai, J. ; Haruta, Yasuhiro ; Uehara, G.
Author_Institution :
Fac. of Pharm. Sci., Hyogo Univ. of Health Sci., Kobe
Volume :
2
Issue :
4
fYear :
2008
Firstpage :
260
Lastpage :
268
Abstract :
Development of drugs requires electrophysiological studies of small animals like mice, rats or guinea pigs. Electrocardiography (ECG) of hirsute animals is time-consuming. We have developed a micro magnetometer array with a 9-channel superconducting quantum interference device (SQUID) with a 2.5-mm diameter pickup-coil for noncontacting measurement of magnetocardiograms (MCGs) in small animals. The micro-MCG successfully recorded the PQRST complex in mice, rats and guinea pigs. A regional myocardial injury was made in rat hearts with a cryoinjury probe, and the characteristic pattern of the injury was recorded in the MCG. An anterior myocardial injury created a QS pattern in the MCG, and a posterior myocardial injury created a QR pattern in the MCG. Quinidine-induced QT prolongation was successfully detected by micro-MCG in mice and rats. Simultaneous recording of ECG and MCG was conducted after intraperitoneal administration of quinidine (60 mg/kg) in guinea pigs. QT interval corrected for heart rate (QTc) in both ECG and MCG correlated well. The newly developed micro-MCG may facilitate electrophysiological studies of small animals, and may enable high-throughput screening of drug-induced QT abnormality.
Keywords :
SQUID magnetometers; bioelectric phenomena; biomedical electrodes; drugs; electrocardiography; magnetocardiography; probes; 9-channel superconducting quantum interference device; ECG recording; MCG recording; QT interval analysis; anterior myocardial injury; cryoinjury probe; drug development; drug-induced QT abnormality screening; electrocardiography; electrophysiological studies; magnetocardiograms; microMCG measurement; micromagnetocardiography system; myocardial injury diagnosis; noncontacting measurement; pickup-coil; posterior myocardial injury; quinidine intraperitoneal administration; quinidine-induced QT prolongation; regional myocardial injury; single-chip SQUID magnetometer array; size 2.5 mm; small animal studies; Animals; Drugs; Electrocardiography; Injuries; Magnetic analysis; Mice; Myocardium; Rats; SQUID magnetometers; Superconducting devices; Electrocardiography; QT prolongation; magnetocardiography; myocardial injury; pharmacological safety test; superconducting quantum interference device;
fLanguage :
English
Journal_Title :
Biomedical Circuits and Systems, IEEE Transactions on
Publisher :
ieee
ISSN :
1932-4545
Type :
jour
DOI :
10.1109/TBCAS.2008.2003979
Filename :
4669630
Link To Document :