• DocumentCode
    984710
  • Title

    Experimental studies of the RF-DC differences of voltage standards

  • Author

    Huang, Dexiang ; He, Shuizhi

  • Author_Institution
    Shaoxing Ind. Res. Inst., Zhejiang, China
  • Volume
    39
  • Issue
    1
  • fYear
    1990
  • fDate
    2/1/1990 12:00:00 AM
  • Firstpage
    10
  • Lastpage
    14
  • Abstract
    The RF-DC differences (d) of primary voltage standards have been determined by experimental and theoretical processes. Comparisons of five different types of voltage standards have been made. The RF-DC differences of some standards at frequencies from 10 Hz to 300 MHz, or even up to 1 GHz, are quite small, but the d value at frequencies from 0.1 to 50 MHz are not as small as they could be. The recommended method for determining d values is to derive an empirical formula and then to correct it from measured data. The uncertainties of primary voltage standards established by this method are ±0.007% at 1 MHz, ±0.01% at 3 MHz, ±0.02% at 10 MHz, ±0.07% at 30 MHz, ±0.15% at 50 MHz, and ±0.3% at 100 MHz
  • Keywords
    measurement standards; voltage measurement; 0.1 to 50 MHz; 1 MHz; 10 Hz to 1 GHz; 10 MHz; 100 MHz; 3 MHz; 30 MHz; RF-DC differences; d value; primary standard; voltage standards; Bolometers; Bridge circuits; Frequency measurement; Frequency response; Helium; Platinum; Semiconductor thin films; Thermistors; Transistors; Voltage;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.50406
  • Filename
    50406