• DocumentCode
    985246
  • Title

    Rectification for cone-beam projection and backprojection

  • Author

    Riddell, Cyril ; Trousset, Yves

  • Author_Institution
    Adv. Med. Applications, GE Healthcare, Buc
  • Volume
    25
  • Issue
    7
  • fYear
    2006
  • fDate
    7/1/2006 12:00:00 AM
  • Firstpage
    950
  • Lastpage
    962
  • Abstract
    The purpose of this paper is to derive a technique for accelerating the computation of cone-beam forward and backward projections that are the basic steps of tomographic reconstruction. The cone-beam geometry of C-arm systems is commonly described with projection matrices. Such matrices provide a continuous framework for analyzing the flow of operations needed to compute backprojection for analytical reconstruction, as well as the combination of forward and backward projections for iterative reconstruction. The proposed rectification technique resamples the original data to planes that are aligned with two of the reconstructed volume main axes, so that the original cone-beam geometry can be replaced by a simpler geometry, where succession of plane magnifications are involved only. Rectification generalizes previous independent results to the cone-beam backprojection of preprocessed data as well as to cone-beam iterative reconstruction. The memory access pattern of simple magnifications provides superior predictability and is, therefore, easier to optimize, independently of the choice of the interpolation technique. Rectification is also shown to provide control over interpolation errors through oversampling, allowing tradeoffs between computation speed and precision to be set. Experimental results are provided for linear and nearest neighbor interpolations, based on simulations, as well as phantom and patient data acquired on a digital C-arm system
  • Keywords
    computerised tomography; image reconstruction; interpolation; iterative methods; medical image processing; phantoms; rectification; C-arm systems; analytical reconstruction; cone-beam forward backprojection; cone-beam forward projection; interpolation errors; iterative reconstruction; linear interpolation; memory access pattern; nearest neighbor interpolation; oversampling; phantoms; rectification; tomographic reconstruction; Angiography; Biomedical equipment; Geometry; Image reconstruction; Interpolation; Medical services; Tomography; X-ray detection; X-ray detectors; X-ray imaging; Cone-beam tomography; iterative reconstruction; rotational angiography; transmission reconstruction;
  • fLanguage
    English
  • Journal_Title
    Medical Imaging, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0278-0062
  • Type

    jour

  • DOI
    10.1109/TMI.2006.876169
  • Filename
    1644810