• DocumentCode
    987696
  • Title

    Thermoelectric Sensors as Microcalorimeter Load

  • Author

    Brunetti, Luciano ; Oberto, Luca ; Vremera, Emil T.

  • Author_Institution
    Ist. Nazionale di Ricerca Metrologica, Turin
  • Volume
    56
  • Issue
    6
  • fYear
    2007
  • Firstpage
    2220
  • Lastpage
    2224
  • Abstract
    Thermoelectric power sensors, instead of bolometers, can now be used as transfer standards in the microcalorimeter technique. This alternative has the technical advantages of being less sensitive to absolute temperature and not downward frequency limited. At the Istituto Nazionale di Ricerca Metrologica, the high-frequency power standards are based on coaxial thermocouples that operate in the frequency range that spans from direct current to 34 GHz. Modified commercial thermocouple mounts in the 7- and 3.5-mm coaxial lines are used to realize the national power standard with an accuracy ranging from 0.03% to 1% in the aforementioned frequency range.
  • Keywords
    calorimeters; calorimetry; measurement standards; power measurement; thermocouples; thermoelectricity; coaxial thermocouples; high-frequency power standards; microcalorimeter load; power sensors; thermoelectric sensors; Bolometers; Calibration; Coaxial components; Frequency; Hafnium; Power measurement; Sensor systems; Temperature sensors; Thermal sensors; Thermoelectricity; Metrology; microcalorimeter; microwaves; power; standards; thermocouples;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.2007.908135
  • Filename
    4389089