• DocumentCode
    989520
  • Title

    CMOS retinal prosthesis with on-chip electrode impedance measurement

  • Author

    Uehara, A. ; Ng, D.C. ; Kagawa, K. ; Tokuda, T. ; Ohta, J. ; Nunoshita, M.

  • Author_Institution
    Nara Inst. of Sci. & Technol., Ikoma Nara, Japan
  • Volume
    40
  • Issue
    10
  • fYear
    2004
  • fDate
    5/13/2004 12:00:00 AM
  • Firstpage
    582
  • Lastpage
    584
  • Abstract
    A retinal prosthesis device with built-in self-test capability is proposed and demonstrated. The measurement of electrode impedance as a self-test is achieved without increasing the chip area by employing analogue multiplexers to allow the electrodes to be used for both stimulation of retinal cells and measurement of impedance. Measurement is performed using a four-terminal method to ensure good accuracy. A prototype stimulus chip with 16×16 channels is fabricated using standard 0.6 μm CMOS technology, and is demonstrated to provide self-test functionality with error of as little as 0.05% in the frequency range of 100 Hz to 100 kHz.
  • Keywords
    CMOS integrated circuits; biomedical electrodes; biomedical electronics; built-in self test; electric impedance; eye; prosthetics; 0.6 micron; 100 Hz to 100 kHz; CMOS retinal prosthesis; CMOS technology; electrodes; four terminal method; on-chip electrode impedance measurement; retinal cells;
  • fLanguage
    English
  • Journal_Title
    Electronics Letters
  • Publisher
    iet
  • ISSN
    0013-5194
  • Type

    jour

  • DOI
    10.1049/el:20040410
  • Filename
    1300217