• DocumentCode
    990794
  • Title

    Implantable 9-Channel Telemetry System for In Vivo Load Measurements With Orthopedic Implants

  • Author

    Graichen, Friedmar ; Arnold, Rüdiger ; Rohlmann, Antonius ; Bergmann, Georg

  • Author_Institution
    Biomech. Lab., Charite - Universitatsmedizin Berlin
  • Volume
    54
  • Issue
    2
  • fYear
    2007
  • Firstpage
    253
  • Lastpage
    261
  • Abstract
    Knowledge of the loads to which orthopedic implants are subjected is a fundamental prerequisite for their optimal biomechanical design, long-term success, and improved rehabilitation outcomes. In vivo load measurements are more accurate than those obtained using mathematical musculoskeletal models. An inductively powered integrated circuit inside the implant measures six load components as well as the temperature and supplied voltage. This low-power circuit includes a 9-channel multiplexer, a programmable memory, a pulse interval modulator, and a radio-frequency transmitter. Together with a few passive components, the integrated circuit is mounted on a ceramic substrate with thick-film hybrid technology. The sensor signals are multiplexed, modulated, and transmitted to an external device. The microcontroller of the external device regulates the alternating magnetic field produced by a power oscillator and synchronizes the pulse interval modulated data stream. A personal computer displays forces, moments, and temperatures in real time. The new telemetry transmitter has, thus far, been used for in vivo load measurements in three patients with shoulder endoprostheses. Eight instrumented vertebral body replacements are ready for implantation, and an instrumented tibial tray is being submitted to laboratory tests
  • Keywords
    BiCMOS integrated circuits; biomechanics; biomedical measurement; biomedical telemetry; orthopaedics; prosthetics; 9-channel multiplexer; ceramic substrate; implantable 9-channel telemetry system; in vivo load measurements; inductively powered integrated circuit; microcontroller; musculoskeletal models; orthopedic implants; power oscillator; programmable memory; pulse interval modulated data stream; pulse interval modulator; radiofrequency transmitter; shoulder endoprostheses; telemetry transmitter; thick-film hybrid technology; vertebral body replacements; Implants; In vivo; Instruments; Integrated circuit measurements; Magnetic field measurement; Orthopedic surgery; Pulse modulation; Radio transmitters; Telemetry; Temperature sensors; Biomedical telemetry; implantable orthopedic device; integrated circuit design; knee; prosthesis; shoulder; spine; tibial tray; Equipment Design; Equipment Failure Analysis; Manometry; Monitoring, Physiologic; Prostheses and Implants; Telemetry; Transducers; Weight-Bearing;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.2006.886857
  • Filename
    4067111