• DocumentCode
    993612
  • Title

    Cryogenic wafer prober for Josephson devices

  • Author

    Geary, J.M. ; Vella-Coleiro, G.P.

  • Author_Institution
    Bell Laboratories, Murray Hill, New Jersey
  • Volume
    19
  • Issue
    3
  • fYear
    1983
  • fDate
    5/1/1983 12:00:00 AM
  • Firstpage
    1190
  • Lastpage
    1192
  • Abstract
    A wafer probing system has been built for the testing of Josephson junction devices at helium temperature. A mechanism moves a probe card from one position to another on a two inch wafer while immersed in liquid helium. The mechanism is actuated by shafts which connect to stepper motors positioned above the helium dewar. A positioning accuracy of ± 50 μ m at the probe tips is achieved. The replaceable probe card is all ceramic and carries 120 rigidly mounted palladium-alloy needles, arranged in signal-ground pairs and positioned in an array which matches the pad design of the particular device under test. Controlled impedance transmission lines are maintained all the way to the wafer´s surface. A computer interface is included so that probing of a whole wafer can be conducted under software control. The system is intended for routine testing of Josephson devices in wafer form as well as for testing very large numbers of individual junctions.
  • Keywords
    Cryogenic materials/devices; Integrated circuit testing; Josephson devices; Ceramics; Cryogenics; Helium; Josephson junctions; Needles; Probes; Shafts; Superconducting devices; System testing; Temperature;
  • fLanguage
    English
  • Journal_Title
    Magnetics, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9464
  • Type

    jour

  • DOI
    10.1109/TMAG.1983.1062335
  • Filename
    1062335