DocumentCode
994004
Title
Capacitive charge division readout of a double-sided germanium strip detector
Author
Inderhees, S.E. ; Kroeger, R.A. ; Johnson, W.N. ; Kinzer, R.L. ; Kurfess, J.D. ; Gehrels, N.
Author_Institution
Universities Space Res. Assoc., Washington, DC, USA
Volume
42
Issue
4
fYear
1995
fDate
8/1/1995 12:00:00 AM
Firstpage
428
Lastpage
431
Abstract
We have implemented a capacitive charge-division readout for a germanium 5×5 orthogonal strip detector. We present measurements of non-linearity, and of energy and position resolution as a function of energy (60 and 662 keV) and of network capacitance. Energy and position non-linearities, which are worse for low capacitance networks, appear in the response along the charge division network, as well as along the length of each detector strip. The nonlinearities can be corrected by using an independent gain calibration for each pixel, allowing for use of a lower capacitance charge division network which provides better position resolution
Keywords
calibration; detector circuits; elemental semiconductors; germanium radiation detectors; nuclear electronics; position sensitive particle detectors; 60 keV; 662 keV; Ge; capacitive charge division readout; double-sided strip detector; energy resolution; independent gain calibration; low capacitance networks; network capacitance; nonlinearity; orthogonal strip detector; pixel detector; position resolution; Capacitance measurement; Capacitors; Detectors; Energy measurement; Energy resolution; Face detection; Germanium; Linearity; Parasitic capacitance; Strips;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.467807
Filename
467807
Link To Document