• DocumentCode
    994371
  • Title

    Increased spatial resolution in transillumination using collimated light

  • Author

    Wist, Abund O. ; Fatouros, Panos P. ; Herr, Steven L.

  • Author_Institution
    Dept. of Radiol., Virginia Commonwealth Univ., Richmond, VA, USA
  • Volume
    12
  • Issue
    4
  • fYear
    1993
  • fDate
    12/1/1993 12:00:00 AM
  • Firstpage
    751
  • Lastpage
    757
  • Abstract
    In traditional transillumination of the breast (diaphanography), the abundance of diffuse light resulting from the use of extended noncollimated sources reduces the visibility of deep seated lesions. A prototype scanning imaging system has been developed to investigate the effectiveness of thin collimated light beams (1.5 mm cross section) synchronized with a similarly collimated detector to increase contrast in lesions normally lost due to the detection of diffuse light. The study demonstrates that detection of opaque 1.5 mm details is possible in phantoms simulating breast tissues 6 mm thick regardless of depth. This is about 10 times better than images obtained on the same samples using present transillumination methods. Furthermore, this study indicates that internal structures (lesions, cysts) in up to 12 mm thick excised breast sections can be visualized by exploiting their frequency-dependent attenuation. This is accomplished by inserting 50 nm interference filters in the input light path, which can be varied in a stepwise manner in the range of 400 nm to 1000 nm. These results demonstrate for the first time that images of lesion-bearing 1 cm or larger tissues can be obtained, thus opening promising possibilities for whole-breast imaging
  • Keywords
    patient diagnosis; 1 cm; 1.5 to 12 mm; 400 to 1000 nm; breast transillumination; collimated light transillumination; cysts; diaphanography; frequency-dependent attenuation; input light path; interference filters; lesions; medical diagnostic technique; spatial resolution; whole-breast imaging; Breast tissue; Frequency synchronization; Imaging phantoms; Interference; Lesions; Optical attenuators; Optical collimators; Prototypes; Spatial resolution; Visualization;
  • fLanguage
    English
  • Journal_Title
    Medical Imaging, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0278-0062
  • Type

    jour

  • DOI
    10.1109/42.251126
  • Filename
    251126