• DocumentCode
    994678
  • Title

    Distributed 3-D iterative reconstruction for quantitative SPECT

  • Author

    Ju, Z.-W. ; Frey, E.C. ; Tsui, B.M.W.

  • Author_Institution
    Dept. of Biomed. Eng., North Carolina Univ., Chapel Hill, NC, USA
  • Volume
    42
  • Issue
    4
  • fYear
    1995
  • fDate
    8/1/1995 12:00:00 AM
  • Firstpage
    1301
  • Lastpage
    1309
  • Abstract
    We describe a distributed three dimensional (3-D) iterative reconstruction library for quantitative single-photon emission computed tomography (SPECT). This library includes 3-D projector-backprojector pairs (PBPs) and distributed 3-D iterative reconstruction algorithms. The 3-D PBPs accurately and efficiently model various combinations of the image degrading factors including attenuation, detector response and scatter response. These PBPs were validated by comparing projection data computed using the projectors with that from direct Monte Carlo (MC) simulations. The distributed 3-D iterative algorithms spread the projection-backprojection operations for all the projection angles over a heterogeneous network of single or multi-processor computers to reduce the reconstruction time. Based on a master/slave paradigm, these distributed algorithms provide dynamic load balancing and fault tolerance. The distributed algorithms were verified by comparing images reconstructed using both the distributed and non-distributed algorithms. Computation times for distributed 3-D reconstructions running on up to 4 identical processors were reduced by a factor approximately 80-90% times the number of the processors participating, compared to those for non-distributed 3-D reconstructions running on a single processor. When combined with faster affordable computers, this library provides an efficient means for implementing accurate reconstruction and compensation methods to improve quality and quantitative accuracy in SPECT images
  • Keywords
    Monte Carlo methods; distributed algorithms; image reconstruction; single photon emission computed tomography; 3D projector-backprojector pairs; SPECT; attenuation; detector response; direct Monte Carlo simulations; distributed 3D iterative reconstruction; distributed algorithms; dynamic load balancing; fault tolerance; heterogeneous network; image degrading factors; master/slave paradigm; multi-processor computers; nondistributed algorithms; projection angles; quantitative SPECT; quantitative single-photon emission computed tomography; reconstruction time; scatter response; single-processor computers; Attenuation; Computed tomography; Degradation; Detectors; Distributed algorithms; Distributed computing; Image reconstruction; Libraries; Reconstruction algorithms; Three dimensional displays;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.467865
  • Filename
    467865