DocumentCode
994767
Title
Direct reconstruction of functional parameters for dynamic SPECT
Author
Limber, M.A. ; Limber, M.N. ; Celler, A. ; Barney, J.S. ; Borwein, J.M.
Author_Institution
Centre for Exp. & Constructive Math., Simon Fraser Univ., Burnaby, BC, Canada
Volume
42
Issue
4
fYear
1995
fDate
8/1/1995 12:00:00 AM
Firstpage
1249
Lastpage
1256
Abstract
We have developed a dynamic SPECT method which allows the prevalent rotating camera systems to be used. In this method, the parameters of the time-activity curve in each pixel are recovered directly from the projection data, without first reconstructing a series of static images. Preliminary test results show that we can accurately recover time-activity curves for a simple model based on activity values expected for myocardial viability studies with fatty acid tracers
Keywords
cardiology; image reconstruction; radioactive tracers; single photon emission computed tomography; direct image reconstruction; dynamic SPECT; fatty acid tracers; functional parameters; myocardial viability; pixel; projection data; rotating camera systems; single photon emission computed tomography; test results; time-activity curve; Biochemistry; Cameras; Head; Image analysis; Image reconstruction; Image resolution; Mathematics; Myocardium; Pixel; Positron emission tomography;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.467872
Filename
467872
Link To Document