Background: Microalbuminuria is a marker for nephropathy in diabetic patients and recognition of
albuminuria is an important step for early detection and evaluation of diabetes complications.
Objective: To assess the reliability of urine dipstick tests for detection of albuminuria.
Methods: The first morning urine samples from candidates for urinanalysis were collected and the
results of dipstick tests for protein on 200 selected samples recorded. Urine albumin concentralion
was determined by electroimmunoassay and the level of urine creatinine determined by Jaffe
method. The albumin/creatinine ratio was further calculated as an index of urine albumin excretion
rate. Comparing with reference method, the sensitivity, specificity, positive predictive value (PV+),
and negative predictive value (PV-) of dipstick test for detection of microalbuminuria were
calculated.
Findings: Based on results obtained by dipstick test, 69 urine samples were negative for presence
of protein, 51 trace (±), 22 (1+),27 (2+), and 31 (3+). Detection limit of dipstick for urine albumin
concentration ranged from 58 to 585 mg/L with a mean of 280 mg/L. When the reference method
for detection of protein in urine samples was used, 35 urine specimens were negative for
albuminuria, 85 with microalbuminuria, and 80 with macroalbuminuria. Dipstick test was found to
produce false positive and negative results of 3 and 25%, respectively. The sensitivity, specificity,
PV+, PV- of dipstick test for detection of microalbuminuria were calculated at 70, 86, 96, and
37.5%, respectively.
Conclusion: Chemical dipstick test is of poor sensitivity for detection of microalbuminuria.
However, the severity of albuminuria among considerable numbers of diabetic patients is high
enough to be detectable by urine dipsticks. While the positive predictive value of chemical urine
dipstick tests are more or less at acceptable limit, the negative predictive values are controversial
and need to be reevaluated by more sensitive and specilic methods
عنوان نشريه :
مجله علمي دانشگاه علوم پزشكي و خدمات بهداشتي - درماني قزوين
عنوان نشريه :
مجله علمي دانشگاه علوم پزشكي و خدمات بهداشتي - درماني قزوين
اطلاعات موجودي :
فصلنامه با شماره پیاپی 49 سال 1387
كلمات كليدي :
#تست#آزمون###امتحان