A Finite-element Analysis of Stress Distribution around Dental Implant in
Different Bar-attachment Systems
پديد آورندگان :
ريسمان چيان، منصور نويسنده دانشكده دندانپزشكي و مركز تحقيقات دندانپزشكي ترابي نژاد، دانشگاه علوم پزشكي اصفهان RISMAN CHIAN, M , دخيل عليان، منصور نويسنده دانشكده دندانپزشكي و مركز تحقيقات دندانپزشكي ترابي نژاد، دانشگاه علوم پزشكي اصفهان DAKHIL ALIAN, M. , قاسمي، احسان نويسنده دانشكده دندانپزشكي و مركز تحقيقات دندانپزشكي ترابي نژاد، دانشگاه علوم پزشكي اصفهان Ghasemi, E , يقيني، جابر نويسنده دانشكده دندانپزشكي و مركز تحقيقات دندانپزشكي ترابي نژاد، دانشگاه علوم پزشكي اصفهان yaghini, jaber
كليدواژه :
ITI (Dolder) and OT barattachment , Implant , Finite-element analysis , باراتچمنت ITI Dolder OT , ايمپلنت , روش المان محدود
چكيده لاتين :
Sta tement of Problems: When implants are used to support an over denture, the optimal stress distribution is
desirable to minimize the forces on the implants.
Purpose: The purpose of this study was to compare stress patterns generated around implant supported over denture
with two different bar attachment systems ITl (Dolder) and OT with finite element method.
Materi als and Method : In this descr iptive study, the stress distr ibution around dental implants in an edentulous
mand ible with two different bar attachment systems lOT without spacer and IT! (Do lder) with spacer) was
calculated by means of three-dimensional, finite-element model of an entire lower jaw. This model was provided
with two endosseo us implants in the canine region. The implants were connected to a 2 mm bar height and a vertical
force was app lied to the first left molar and gradually increase d from a to SON. Then, resultant stress distribution
was eva luated.
Result: In OT and IʹTl (Dolder) bar attachments, the maximum von Mises stress values were observed in the
cortical bone adj acent to the implant neck and this value was higher for OT bar (6.1l4M pa) than ITI (Do lder) bar
attachment (3.896 Mpa).
Conclusion: This 3-D FE study suggested that the use of ITI (Do lder) bar attachment with spacer is more
adva ntageo us than OT bar attachment for implant-supported overdenture with regard to optimizing stress
distribution.