عنوان مقاله :
تاثير مولفه هاي هيدرودايناميك در كاهش ياخته هاي اندوتليوم قرنيه بعد از فيكوامولسيفيكيشن
عنوان به زبان ديگر :
Corneal Endothelial Cell Loss after Phacoemulsification
in Relation to Hydrodynamic Parameters:
Low versus High Vacuum Techniques
پديد آورندگان :
رحمتي كامل، محسن نويسنده مركز چشم پزشكي هلال ايران RAHMATI KAMEL, M. , برادران رفيعي، علي رضا نويسنده دانشگاه علوم پزشكي شهيد بهشتي BARADARAN RAFIEI, A.R. , كياوش، ويكتوريا نويسنده KIAVASH, V. , عين الهي ، بهرام نويسنده دانشگاه علوم پزشكي شهيد بهشتي؛معاون آموزشي وزارت متبوع EINOLLAHI, B. , كريميان، فريد نويسنده دانشگاه علوم پزشكي شهيد بهشتي KARIMIAN, F. , ميردهقان، علي نويسنده دانشگاه علوم پزشكي شهيد بهشتي MIRDEHGHAN, A. , زارع، محمد نويسنده دانشگاه علوم پزشكي شهيد بهشتي ZARE, M.
اطلاعات موجودي :
فصلنامه سال 1388 شماره 56
كليدواژه :
ياخته هاي اندوتليوم قرنيه , فيكوامولسيفيكيشن , واكيوم و فلوريت , هيدرودايناميك
چكيده لاتين :
Purpose: To evaluate the effect of vacuum and flow rate on endothelial cell loss after high versus low
vacuum phacoernulsification.
Methods: This randomized clinical trial was performed on 60 eyes of 60 patients with moderate lens opacity
(nuclear sclerosis 3+). All surgeries were performed by one experienced surgeon using stop and chop
technique with Sovereign white star machine (AMO). Patients were randomly assigned to high and low
vacuum techniques in equal numbers. The machine was set on 400 mmHg vacuum and 40 ml/min flow rate
in the high vacuum group and on 200 mmHg vacuum and 20 ml/min flow rate in the low vacuum group
during the chop stage. All other parameters were similar in both groups. Phacotime multiplied by average
ultrasound power was defined as total ultrasound energy. Specular microscopy was performed before and
I, 6 and 12 weeks after the opera tion.
Results: After 12 weeks, mean endothelial cell loss was 9.0±4.0% versus 9.6±4.6% in the low and high
vacuum groups, respectively (p~0.6). Mean ultrasound power was 9.2±4.3% and 13.1±4.6% in the low and
high vacuum groups, respectively (p~0.001). Mean phacotime was 1.28±1.0 minutes in the low vacuum
group versus 0.88±0.6 minutes in the high vacuum group (P~0.04). Total ultrasound energy and total fluid
volume used (turbulence) during phacoemulsification was similar between the two study groups. Total
ultrasound energy was the most powerful predictor of endothelial cell loss (Rʹ~0936, P~()ʹ()Ol), but
turbulence was not a significant predictor (R2~0.924, P~O.l).
Conclusion: No significant difference in endothelial cell loss was found between low and high vacuum
techniques. This study supports advice to junior surgeons to choose lower hydrodynamic phaco machine
parameters; experienced surgeons can choose higher parameters to reduce phacotirne