Depth dose of critical organs of phantom based on surface
dose exposed with dual X-ray absorptiometry: pencil
beam using TLD dosimetry
پديد آورندگان :
شرفي، علي اكبر نويسنده گروه راديولوژي،دانشگاه علوم پزشكي ايران SHARAFI, A.A. , مختاري ديزجي، منيژه نويسنده دانشكده فني و مهندسي- دانشگاه تربيت مدرس MOKHTARI DIZJI, M. , مخلصيان، نرگس نويسنده دانشگاه آزاداسلامي،واحد علوم وتحقيقات MOKHLESIAN, N , لاريجاني، باقر نويسنده مركز تحقيقات غدد درون ريز و متابوليسم -دانشگاه علوم پزشكي تهران Larijani, B , كاظم نژاد، انوشيروان نويسنده دانشكده علوم پزشكي-دانشگاه تربيت مدرس Kazemnezhad, A
اطلاعات موجودي :
فصلنامه سال 1387 شماره 30
كليدواژه :
Depth dose , Critical organs , اندام هاي بحراني , Dual X-ray absorptiometry , dosimetry , جذب سنجي دوگانه اشعه X , دزيمتري , دز عمقي , فانتوم , Phantom
چكيده لاتين :
Introduction: Dual X-ray absorptiometry (DXA) is one of the most widely used techniques for
non-invasive assessment of bone status, Radiation dosimetry is well established technique for pencil
beam and fan beam DXA system, for the assessment of the surface absorbed dose. No published
assessment of the absorbed dose for the various depths of the critical organs such as the thyroid and
uterus WaS found. Therefore, in this study, we measured the surface dose and depth dose of critical
organs to determine the correlation between the depth dose and the surface dose.
Materials and Methods: A Lunar DPX-MD (pencil beam) system was used in this study. An
anthropomorphic phantom WaS designed, AP spine and femur scan modes were used to measure the
surface and depth doses of the thyroid left and right lobes and uterus in various deeps and Scan centers.
TLDs-400 were placed at the surface, near the source and also inserted at different depths in thyroid
and uterus of the anthropomorphic phantom, Absorbed doses were measured on the phantom for AP
spine and femur scans, The correlation between the absorbed dose and the depth was found using the
linear regression analysis.
Results: There was no significant correlation between the depth dose and the SCan center dose
except in the femur scan. AP spine effective dose were calculated as 0.064, 0.059, 0.061 and 0.242
~Sv for thyroid left, right lobes, uterus and ovary, respectively.
Conclusion: It is concluded that there is significant correlation between the surface and the dose
of various depths of the scanned sections