عنوان مقاله :
بررسي عملكرد دستگاه هاي سترون كننده بيمارستاني با روش بيولوژيك و عوامل مؤثر بر آن
عنوان به زبان ديگر :
Assessment of hospital sterilizer equipments performance and related effective factors
پديد آورندگان :
دبيران ، سهيلا نويسنده dabiran, soheyla , رضوي ، منصور نويسنده razavi, mansour , ميرافشار، محمد نويسنده دانشگاه علوم پزشكي تهران mir afshar, mohammad
اطلاعات موجودي :
فصلنامه سال 1388
كليدواژه :
استريليزاسيون , CSR , اتوكلاو بخار , فور , اتيلن اكسايد
چكيده لاتين :
sterilizer devices , that are presented in affiliated hospitals to Tehran University of medical sciences with biologic monitoring way .
Methods: This is a cross - sectional study conducted in all 135 devices that are presented in CSR units of 14 hospitals and related responsible staff. In this study, sterilization process was assessed by using of spores, microbial vials and culturing of the samples and also, the Knowledge of CSR personnel was measured by a valid and reliable questionnaire. We have used , T-Test for data analysis in this study.
Results: 14 out Of 135 devices (10.4 %) were not in good function . there was a significant relationship between age of personnel and their knowledge. There were significant relations between the sex (0.001) , educational grade (0.02) , years of experiences , and the knowledge of the staff (P=0.001) and also CSR unit area in this study.
Conclusion: this study suggests that hospitals must have CSR units with sufficient area and educated personals in related fields. They must also regularly educate the staff , especially males that are responsible in CSR units.
• ORIGINAL ARTICLE CQDE:05
Assessment of hospital sterilizer equipments performance and related effective factors
Key words: Sterilization , CSR , Steam Autoclave , Oven , Ethylene Oxide